S12 Vet Monitor User’s Manual
59
D.2.11 System Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
Battery Low High Not clearable
Please connect AC power supply for
power supply and charge the battery.
Battery Fault Prompt / Please replace the battery.
Recorder No Paper Low
Sound and light
can be cleared
The recorder is not loaded with paper
or the recorder door is not closed.
Please check the recorder to make sure
the paper is loaded or the recorder door
is closed.
Recorder Not Exist Prompt /
The recorder module is not plugged in.
please insert the recorder module.
Recorder Too Hot Low Not clearable
The recorder works too long. Please
restart the recording task after the
recorder head cools down.
CMS Disconnected Med
Completely
clear
The monitor is disconnected from the
CMS. Please check the network
connection.
Disk Full Med Not clearable
The storage space of the monitor is full.
Please clear the patient related data in
time.
Disk Will Be Full Low
Sound and light
can be cleared
The storage space of the monitor is
almost full. Please clear the patient
related data.