S12 Vet Monitor User’s Manual
7-13
can output a signal to the nurse call system, call the nurse. The monitor provides a nurse
call connector, and the monitor is connected to the nurse call system of the hospital
through the randomly provided nurse call cable. After the system is connected, the
connector can implement the nurse call function.
The nurse call function must be valid only if the following conditions are met:
The nurse call function is turned on.
A user-defined alarm occurs.
The monitor is not alarm paused or off.
7.10.1. Changing Nurse Call Settings
To set the type and priority of alarms that are sent to the nurse call system, follow this
procedure:
1. Select Main Menu quick key→from System column to select
Maintenance→input maintenance password→Enter.
2. SelectAlarmsubmenu→Nurse Callsubmenu.
3. SelectSignal Typeto set the type of nurse call signal.
Pulse: The nurse call signal is a pulse signal and each pulse lasts one
second. When multiple alarms simultaneously occur, only one pulse
signal is outputted. If an alarm occurs but the previous one is not cleared,
a new pulse signal will also be outputted.
Continuous: The nurse call signal lasts until the alarm ends. That is
to say the duration of a nurse call signal is equal to that of the alarm
condition
.
4. Select Trigger Typeto set the work mode of the nurse call relay.
5. Select Alarm Priorityto set the priority of alarms sent to the nurse call
system.
6. Select Alarm Typeto set the type of alarms sent to the nurse call system.
WARNING
Do not rely exclusively on the nurse call system for alarm notification.