Query Syntax OCP:IEND?
Query Response <NR2>
5.8 OCP:STEP
This command is used to set the current step of OCP test.
Command Syntax OCP:STEP<NR1>
Parameters 1 1000
Example OCP:STEP 500
Query Syntax OCP:STEP?
Query Response <NR2>
5.9 OCP:DWELl <NRf+>
This command is used to set the single-step dwell time of OCP test.
Command Syntax OCP:DWELl <NRf+>
Parameters 0.00001 0.99999
Unit
S
Example OCP:DWEL 0.01 or OCP:DWEL 10ms
Query Syntax OCP:DWELl?
Query Response <NR2>
5.10 OCP:VTRig <NRf+>
This command is used to set the OCP trigger level.
Command Syntax OCP:VTRig <NRf+>
Parameters MIN MAX
Unit
V
Example OCP:VTR 11.8
Query Syntax OCP:VTRig?
Query Response <NR2>
5.11 OCP:RESult[:OCP]?
This command is used to query the current value of OCP.
Command Syntax OCP:RESult[:OCP]?