Query Syntax [SOURce:]DYNamic:LOW:DWELl?
Query Response <NR2>
*RST Value
0.00002
7.23 [SOURce:]DYNamic:SLEW <NRf+>
This command is used to set the current slope in dynamic mode.
Command Syntax [SOURce:]DYNamic:SLEW <NRf+>
Parameters MIN MAX- MINimum- MAXimum
Unit
A/uS
*RST Value
MAX
Example DYN:SLEW 3
Query Syntax [SOURce:]DYNamic:SLEW?
Query Response <NR2>
7.24 [SOURce:]DYNamic:SLEW:RISE <NRf+>
This command is used to set the current rising slope in dynamic mode.
Command Syntax [SOURce:]DYNamic:SLEW:RISE <NRf+>
Parameters MIN MAX- MINimum- MAXimum
Unit
A/uS
*RST Value
MAX
Example DYN:SLEW:RISE 3
Query Syntax [SOURce:]DYNamic:SLEW:RISE?
Query Response <NR2>
7.25 [SOURce:]DYNamic:SLEW:FALL <NRf+>
This command is used to set the current falling slope in dynamic mode.
Command Syntax [SOURce:]DYNamic:SLEW:FALL <NRf+>
Parameters MIN MAX- MINimum- MAXimum
Unit
A/uS
*RST Value
MAX