EasyManua.ls Logo

BK Precision 8500B Series - 4.3 Input Connection and Wiring; 4.3.1 Current Carrying Capacity

BK Precision 8500B Series
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
MAXimum(high range)
Example SOUR:VOLT:RANG MIN
Query Syntax [SOURce:]VOLTage:RANGe?
Query Response <NR2>
7.5 [SOURce:]VOLTage:RANGe:AUTO[:STATe]
This command is used to set the auto-range state of the voltage meter of the load module.
Command Syntax [SOURce:]VOLTage:RANGe:AUTO[:STATe] <bool>
Parameters 0 - 1 - ON - OFF
Query Response 1
Example VOLT:RANG:AUTO 1
Query Syntax [SOURce:]VOLTage:RANGe:AUTO[:STATe]?
Query Response <NR1>
7.6 [SOURce:]VOLTage:[LEVel:]ON <NRf+>
This command sets the voltage of sink current on.
Command Syntax [SOURce:]VOLTage[:LEVel]:ON <NRf+>
Parameters None
Unit
V
*RST Value
1
Example VOLT:ON 3
Query Syntax [SOURce:]VOLTage[:LEVel]:ON?
Query Response <NR2
7.7 [SOURce:]Voltage:[LEVel:]OFF <NRf+>
This command is used to set the load start unloading voltage value.
Command Syntax [SOURce:]VOLTage[:LEVel]:OFF <NRf+>
Parameters MIN MAX - MINimum - MAXimum
Unit
V
*RST Value
0.5
Example VOLT:OFF 2

Table of Contents

Other manuals for BK Precision 8500B Series

Related product manuals