12.50 Read the max measured current in list mode. (A4H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A4H
4-7 max measured current (Low byte to High byte order) value resets after query
8-25 System reserve
26 Checksum
12.51 Read the min measured current in list mode. (A5H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A5H
4-7 min measured current (Low byte to High byte order) value resets after query
8-25 System reserve
26 Checksum
12.52 Read on-load capacitance. (A6H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A6H
4-7 On-load capacitace (Low byte to High byte order) value resets after query
8-25 System reserve
26 Checksum
12.53 Set/Read current rising slope. (B0H/B1H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command B0/B1H
4-7 Current rising slope (Low byte to High byte order)
8-25 System reserve
26 Checksum