*RST Value
MAXimum
Example CURR:PROT 3
Query Syntax [SOURce:]CURRent:PROTection[:LEVel]?
Query Response NR2
7.14 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
This command is used to set the current setting in CC mode.
Command Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters MIN MAX- MINimum- MAXimum
Unit
A
*RST Value
MINimum
Example CURR 5
Query Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Query Response <NR2>
7.15 [SOURce:]POWer:PROTection[:LEVel] <NRf+>
This command sets the power protection level. If the input power exceeds the power protection level for the time specied
by POW:PROT:DEL, the input is turned o.
NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions caused by programmed changes from tripping
the over power protection.
Command Syntax [SOURce:]POWer:PROTection[:LEVel] <NRf+>
Parameters MIN MAX - MINimum - MAXimum
Unit
W
*RST Value
MAXimum(high range)
Example POW:PROT 100
Query Syntax [SOURce:]POWer:PROTection[:LEVel]?
Query Response <NR2>
7.16 [SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRf+>
This command is used to set the power setting in CP mode.
Command Syntax [SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters MIN MAX- MINimum- MAXimum

Table of Contents

Other manuals for BK Precision 8500B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 8500B Series and is the answer not in the manual?

BK Precision 8500B Series Specifications

General IconGeneral
BrandBK Precision
Model8500B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals