12.19 Save/Recall list le in appointed area.. (4CH/4DH)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command 4CH/4DH
4 Storing area (1-7)
5-25 Reserved
26 Checksum
12.20 Setting/Reading timer value of FOR LOAD ON (50H/51H
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command 50H/51H
4 The lowest byte of time value in timer. (1 represent 1S)
5 The highest byte of time value in timer.
8-25 Reserved
26 Checksum
Time unit in Timer is S, 1S is represented by 1.
12.21 Disable/Enable timer of FOR LOAD ON (52H); Read timer state of FOR LOAD ON (53H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command 52H/53H
4 Timer state (0:OFF, 1:ON)
5-25 Reserved
26 Checksum
12.22 Set communication address (54H)
Byte Description
1 AAh
2 Address (0-0XFE)
3 Command 54H
4 New communication address (0-31)
5-25 Reserved
26 Checksum