35
Voltage Rise/Fall Time Test
9.1 TIME:VOLTage:LOW <Voltage> 35
9.2 TIME:VOLTage:HIGH <Voltage> 35
9.3 TIME:VOLTage:UP? 35
9.4 TIME:VOLTage:DOWN? 35
9.1 TIME:VOLTage:LOW <Voltage>
This command is used to set the start rising voltage value of the load when measuring the rise time of voltage output.
Command Syntax TIME:VOLTage:LOW <Voltage>
Parameters 0 through MAX- MINimum- MAXimum- DEFault
Unit
V
Query Response MINimum
Query Syntax TIME:VOLTage:LOW?
9.2 TIME:VOLTage:HIGH <Voltage>
This command is used to set the starting drop voltage value of the load when measuring the fall time of voltage output.
Command Syntax TIME:VOLTage:HIGH <Voltage>
Parameters 0 through MAX- MINimum- MAXimum- DEFault
Voltage Rise/Fall Time Test
Unit
V
Query Response MAXimum
Query Syntax TIME:VOLTage:HIGH?
9.3 TIME:VOLTage:UP?
This command is used to read the voltage rise time.
Command Syntax TIME:VOLTage:UP?
Parameters None
Query Response <NRf>
9.4 TIME:VOLTage:DOWN?
This command is used to read the voltage fall time.
Command Syntax TIME:VOLTage:DOWN?
Parameters None
Query Response <NRf>