12.46 Read the information of load (on-load capacitance, on-load time). (A0H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A0H
4-7 On-load capacitance (Low byte to High byte order)
8-11 On-load time or rising/falling slope (Low byte to High byte order)
12-15 The rest time of timer (Low byte to High byte order)
16-25 Reserved
26 Checksum
12.47 Read the information of E-load (max/min input voltage/current). (A1H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A1H
4-7 Max input voltage (Low byte to High byte order)
8-11 Min input voltage (Low byte to High byte order)
12-15 Max input current (Low byte to High byte order)
16-19 Min input current (Low byte to High byte order)
20-25 Reserved
26 Checksum
12.48 Read the max measured voltage in list mode. (A2H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A2H
4-7 Max measured voltage (Low byte to High byte order) value resets after query
8-25 System reserve
26 Checksum
12.49 Read the min measured voltage in list mode. (A3H)
Byte Description
1 AAh
2 Address (0-31, 0XFF)
3 Command A3H
4-7 Min measured voltage (Low byte to High byte order) value resets after query
8-25 System reserve
26 Checksum