7 Preparation
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 107
7. Preparation
7.1. Preparation of the accessories
Prepare the accessories (e.g. surgical handles, instruments, active
electrodes, neutral electrodes and cables) as described in the
corresponding operating manuals.
Check the accessories before and after use for damage and to ensure
that they are working properly.
7.2. Disinfection and cleaning
NOTE
Incorrect handling of the HF device can cause damage to the unit!
Never sterilize the ARC 400 device. Instead, clean or disinfect it.
WARNING
Risk of electric shock and fire!
Unplug the power connection before cleaning the device.
For cleaning surfaces, use the approved cleaning
agents/disinfectants only as specified by the manufacturer.
Ensure that no liquid penetrates the device.
Ensure that the AUTOSTART function is deactivated.
1. Apply the cleaning agent and disinfectant.
BOWA recommends the use of cleaning and disinfection agents which
are suitable for surface cleaning of medical devices made of plastic,
metal and glass.
The manufacturer accepts no responsibility if other types of cleaning
and disinfecting agents are used.
Follow the instructions provided by the manufacturer of the cleaning
agent.
2. Wipe the agent off with a sponge moistened with clean water or with a
cloth.
3. Dry the device using a clean, lint-free cloth.