12 EMC
160 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
12. EMC
Medical electrical equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the
EMC information provided in this document.
IEC 60601-1-2; Ch. 5.2.2.1a)
Portable and mobile RF communication equipment can affect medical
electrical equipment (see also table 4 and 6 on pages 161 and 162)
IEC 60601-1-2; Ch. 5.2.1.1b)
The generator should not be used adjacent to or stacked with other
electrical equipment. If adjacent or stacked use is necessary both the
generator and other equipment should be observed to verify normal
operation in the configuration in which it will be used.
IEC 60601-1-2; Ch. 5.2.2.1d)
The use of accessories and cables other than those specified may
result in increased emissions or decreased immunity.
IEC 60601-1-2; Ch. 5.2.2.1b)
This HF device is solely intended to be used by trained medical
personnel. This HF device may cause functional impairment or
interference to the operation of other devices in the vicinity. It may be
necessary to take suitable remedial measures, such as changing the
orientation, arrangement or screening of the HF device.
IEC 60601-1-2:2007; Section 5.2.1.4
12.1. Guidelines and manufacturer’s declaration
in accordance with IEC 60601-1-2:2007
Emission of electromagnetic interference (IEC 60601-1-2, Table 1)
The ARC 400 is intended for operation in an electromagnetic environment as described below. The customer or
user of the ARC 400 should ensure that it is operated in such an environment.
Interference emission measurement
Conformity
Electromagnetic environment
guideline
HF emissions according to CISPR 11
Group 2
The ARC 400 must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic devices may be affected.
HF emissions according to CISPR 11
Class A
The ARC 400 is suitable for use in
facilities other than those suitable for
a residential environment or those
connected directly to the public
power grid, which also supplies
power to buildings used for
residential purposes.
Emission of harmonics according to
IEC 61000-3-2
Classes A
Emission of voltage fluctuations and
flicker according to IEC 61000-3-3
Conforms