5 Operation
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 59
5.6.3. Dry
This mode is used for monopolar dry cutting. A large, controlled arc is
generated, which allows significantly deeper coagulation to be obtained.
Application areas
Cardiac surgery and blood coagulation in retracting blood vessels in the
sternum region.
Suitable instruments
Knife electrodes
5.6.4. Argon
This mode is used to perform open surgical interventions in combination
with the ARC PLUS companion device for argon-assisted cutting. With
suitable instruments connected, argon-assisted cutting can be performed
using rigid electrodes.
Application areas
Visceral surgery
Suitable instruments
Rigid argon electrodes
Argon handle