5 Operation
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 79
5.9.11. Vaporisation (optional)
This mode is used for bipolar blood coagulation in gynaecology and in
urology for vaporisation.
Be sure to use NaCl as the rinsing liquid.
Perform continuous rinsing during the application.
Use only conductive gel to avoid damage to the urinary tubes.
Avoid continuous activation.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment
of bladder tumors (TUR-B), vaporization of prostate tissue (TUR-VAP)
Suitable instruments
Resectoscope
Rollerblade electrode
Vaporisation electrode
Make sure that the instrument has contact with the tissue while activating
bipolar coagulation to avoid an unintended heating of the irrigation fluid.
This function is available if the device has the option Bipolar Resection
(900-395).
5.9.12. Bipolar SimCoag (optional)
This mode is used for coagulation when using two bipolar instruments, e.g.
forceps. The power is individually selectable for each instrument without
any loss in power output during simultaneous activation.
The powercan be set in steps of 5 watt.
Application areas
Simultaneous coagulation and preparation with two bipolar instruments in
general surgery, vascular surgery, plastic surgery, traumatology, Neuro
surgery and Orthopedics.
Suitable instruments
Bipolar forceps
Bipolar scissors
This function is available if the device has the option Bipolar SimCoag
(900-399).