5 Operation
78 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
5.9.10. Bipolar resection (optional)
This mode is used for bipolar blood coagulation in gynaecology and in
urology for resection under conductive rinsing liquid (saline solution).
Be sure to use NaCl as the rinsing liquid.
Perform continuous rinsing during the application.
Use only conductive gel to avoid damage to the urinary tubes.
Avoid continuous activation.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment
of bladder tumors (TUR-B), vaporization of prostate tissue (TUR-VAP)
Suitable instruments
Resectoscope
Resection sling
Rollerblade electrode
This function is available if the device has the option Bipolar Resection
(900-395).
Make sure that the instrument has contact with the tissue while activating
bipolar coagulation to avoid an unintended heating of the irrigation fluid.