EasyManua.ls Logo

Bowa ARC 400 - Tissueseal PLUS (Optional); Bipolar Scissors

Bowa ARC 400
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Operation
76 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
5.9.6. TissueSeal PLUS (optional)
This mode is used for the permanent sealing of veins, arteries and tissue
bundles with TissueSeal PLUS
®
for open surgical applications.
Conventional instruments cannot be used in this mode. The output power is
preconfigured and automatically regulated.
Application areas
Vessel sealing in open surgery
Suitable instrument
TissueSeal PLUS
®
This function is available if the device has the option LIGATION
(900-396).
5.9.7. Bipolar scissors
This mode is used with bipolar scissors. It can be used for coagulation
before or during cutting, point coagulation, coagulation of cuts and surface
coagulation.
Application areas
Preparation, coagulation and cutting of tissue
Suitable instruments
Bipolar scissors
Bipolar scissors should only be operated with the current type bipolar
scissors cutting or bipolar scissors coagulation.