4 Prearrangement
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 37
4.6. Turning the device off
Press the On/Off button (1) to switch off the device. Also switch off the
power switch (31) on the rear panel (mains disconnection).
NOTE
Switching off the device with the front panel On/Off button
(Suspend Mode)
After the medical device has been switched off and on by the
On/Off button 1 on the front panel, the last (most recently) saved
parameters (defined settings) are reloaded.
In an emergency, an enabled output can be disabled at any time
by switching off the On/Off button 1 on the front panel.
NOTE
Switching off the device with the power switch on the rear panel
(power interruption longer than 15 seconds)
After a supply voltage interruption longer than 15 seconds, the
parameter settings for the currently selected program that were last
saved in non-volatile memory are reloaded.
NOTE
Brief power interruption shorter than 15 seconds
After a supply voltage interruption shorter than 15 seconds, the
parameter settings for the currently selected program that were last
saved in volatile memory are reloaded.