6 Detecting and correcting faults
106 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
6.2. Fault indications for EASY monitoring
Fault indications are displayed in three stages (green, yellow and red)
when problems occur.
When working with a split neutral electrode, the following faults may occur:
EASY
monitoring
Cause
Indication
Corrective measures
Flashes yellow
Significant increase in
resistance
Depending on the
indication, there may be
heating under the neutral
electrode
Stopping the application is
not necessary.
Check the proper
application of the
neutral electrode.
Switches from
green to
continuous red
A significant problem
occurred when the
monopolar current was
activated
An acoustic
signal sounds.
A warning
message
appears on the
display
Notice
(grey screen)
Warning
(orange
screen)
Error
(red screen)
Check the neutral
electrode and neutral
electrode cable (see
Section EASY neutral
electrode monitoring
(EASY monitoring),
page 35.
Check the neutral
electrode cable for
proper connection and
external damage.
Loosened electrode
An acoustic
signal sounds.
A warning
message
appears on the
display
Notice
(grey screen)
Warning
(orange
screen)
Error
(red screen)
Reattach the neutral
electrode. If the fault
persists, replace the
neutral electrode.