BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 13
2.3. Personal safety instructions
2.3.1. Ambient conditions
Do not use the HF device in the immediate vicinity of the patient.
Observe the minimum distances recommended by BOWA, as shown
in the following figure.
2.3.2. Patients with pacemakers
Malfunction or destruction of the pacemaker can endanger the life of the
patient or result in irreversible injuries to the patient.
In the case of patients with pacemakers, consult the cardiologist
before carrying out HF surgery.
Use bipolar HF methods.
Attach the HF neutral electrode close to the operating field.
Set the demand pacemaker to a fixed frequency.
Ensure that the pacemaker does not come into contact with the HF
electrode.
Keep a fully operational defibrillator within reach.
Carry out a postoperative pacemaker check.