5 Operation
62 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
5.6.9. GastroLOOP 2
This mode is used in gastroenterology. Polypectomy snares are used for
cutting and coagulation. ARC control generates the cutting effect with
simultaneously minimized output power. This mode consists of a series of
cutting current pulses followed by a coagulation phase. With an accelerated
pulse rate of 1.5 cutting pulses per second, this mode is suitable for
experienced users.
Application areas
Polyp removal using polypectomy snares and flexible endoscopy, with
accelerated pulse rate for experienced users.
Suitable instruments
Polypectomy snares
5.6.10. GastroLOOP 3
This mode is used in gastroenterology. Polypectomy snares are used for
cutting and coagulation. ARC control generates the cutting effect with
simultaneously minimized output power. This mode consists of a series of
cutting current pulses followed by a coagulation phase. With an accelerated
fast pulse rate of 2.2 cutting pulses per second, this mode is suitable for
advanced users.
Application areas
Polyp removal using polypectomy snares and flexible endoscopy, with
accelerated fast pulse rate for advanced users.
Suitable instruments
Polypectomy snares