10 Technical specifications
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 115
Mode icon
Description
CCS
ARC CONTROL
Form of HF voltage
Max. power
output
Peak voltage
Default
values
Rated load resistance
Modulation
Effect
Power range
Effect
Max. Watt
Frequency
Duty cycle
Modes Monopolar Cutting
Standard
Yes
Yes
sinusoidal
constant
1
2
3
4
5
6
7
8
9
1 W
400 W
400 Vp
450 Vp
560 Vp
650 Vp
650 Vp
700 Vp
700 Vp
700 Vp
750 Vp
5
100
200 Ω
---
---
Micro
Yes
yes
sinusoidal
constant
1
2
3
4
5
6
7
8
9
1 W
50 W
280 Vp
340 Vp
380 Vp
400 Vp
400 Vp
400 Vp
450 Vp
450 Vp
450 Vp
5
20
500 Ω
---
---
Dry
Yes
yes
sinusoidal
modulated
1
2
3
4
5
6
7
8
9
1 W
200 W
1.4 kVp
1.4 kVp
1.4 kVp
1.4 kVp
1.5 kVp
1.6 kVp
1.6 kVp
1.6 kVp
1.6 kVp
5
100
500 Ω
20 kHz
30 %
30 %
30 %
30 %
25 %
20 %
20 %
20 %
20 %
Argon
Yes
yes
sinusoidal
constant
1
2
3
4
5
6
7
8
9
1 W
300 W
400 Vp
450 Vp
560 Vp
650 Vp
650 Vp
700 Vp
700 Vp
700 Vp
750 Vp
5
100
500 Ω
---
---
Resection
Yes
yes
sinusoidal
constant
1
2
3
4
5
250 W
650 Vp
700 Vp
700 Vp
700 Vp
750 Vp
2
---
500 Ω
---
---
MetraLOOP
Yes
Yes
sinusoidal
constant
1
2
3
300 W
350 W
400 W
650 Vp
1
---
100 Ω
---
---
Laparoscopy
Yes
Yes
sinusoidal
constant
1
2
3
4
5
6
7
8
9
1 W - 200W
400 Vp
450 Vp
560 Vp
650 Vp
650 Vp
700 Vp
700 Vp
700 Vp
750 Vp
5
100
500 Ω
---
---