10 Technical specifications
140 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
Monopolar Coagulation Cardiac Thorax
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Cardiac Thorax"
= 50 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Cardiac Thorax"
= 100 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Coagulation Cardiac Thorax"
Rated load resistance = 500
HF output voltage U [Vp] for the setting
"Monopolar Coagulation Cardiac
Thorax" (idle mode)
= 1800 Vp