Disassembly / Assembly SpaceStation
3
SpaceStation, 1.0 gb 3- 25
1.0
Check List for Checks after Repair
Visual Inspection Electrical Safety
according to IEC / EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection
Cleanliness
Completeness
Damage and faults affecting safety
Damage to and readability of the label
Cover caps
Connectors
Well running and sealing of the module
lock
Mains cable
Accessories
Mains voltage according to the TSC
Protective conductor resistance ac-
cording to the TSC
Earth leakage current according to the
TSC
Locking with second SPS
Pump locking of all slots
Connect unit to mains.
General functional check with SpaceCover
comfort and pump of every slot.
Voltage supply of all slots
Alarm transmission
Staff call transmission