52
10.3 Calibration Setting
10.3.1 Selecting the Calibration Type
Set the calibration type that will be used in Calibration Execution.
To set up “Span calibration using the built-in weight”,(UW only) Select menu item 1 .
To set up “Calibration check using the built-in weight”,(UW only) Select menu item 2 .
To set up “Span calibration using external weights”, Select menu item 3 .
To set up “Calibration check using external weights”, Select menu item 4 .
10.3.2 PSC Fully-automatic Calibration (UW series only)
With the PSC function, span calibration is performed automatically using the built-in calibration
weight when the balance detects a temperature change that would affect weighing accuracy.
1. To turn ON the PSC function, Select menu item
5 .
2. To turn OFF the PSC function, Select menu
item 6 .)
x Blinking calibration symbol indicates an automatic calibration is about to start.
x If PSC starts while the balance is in use, press the [POWER] key to abort that cycle.