14.3 Peak Hold
Detects the peak value of a fluctuating weight. “Peak value” is the highest or lowest value
displayed in the duration after it has changed beyond five times the Zero Range until the
first stability detection after that. The “P” symbol (“P” of Auto-Print symbol) is illuminated
when the Peak Hold function is activated. Select menu item 49 in advance.
1. In the peak detection standby state with the “P”
symbol and the Stand-by symbols illuminated,
press the [O/T] key to tare the display.
2. Press the [PRINT] key.
The Stand-by symbol disappears and peak value
detection starts.
3. Upon stability is detected, the operation of peak
value detection is ceased.
The detected peak value is displayed with the “P”
symbol and Asterisk simultaneously illuminated.
The peak value is output to external devices if
connected. After the detection of the peak value,
the same display remains regardless of the load
change on the pan.
4. Press the [POWER] key for the next operation.
The balance returns to the peak detection
standby state described in the above step 1.