11. Environment
11.1 Overview
Settings on the balance can be changed to compensate for the installation environment such as the
degree of vibration or air movement or for the purpose of weighing a solid, liquid or powder.
11.2 Stability and Response (Averaging)
It is possible to match the stability of the display and the degree of response with the requirements of
specific applications or the installation environment. One of the five modes can be selected. Note
that adjustments for stability and response conflict with each other, although the UW/UX series is
designed to meet both.
Auto mode: Select menu item 22 .
The balance automatically performs optimum averaging dynamically while
observing the load data. This is the recommended setting and should be used
unless special circumstances exist.
Pouring mode: Select menu item 23 .
(Filling, dosing) This mode is particularly suitable for dosing or filling purposes.
Note that this mode is very sensitive to wind and vibration.
(Environmental setting in Pouring mode)
Pouring mode allows further adjustments to the surrounding environmental
conditions. Choose the optimal setting for your installation site observing the
response and stability.
Every time the circle in the right part of the [O/T] key is pressed during
weighing in Pouring mode, the environmental setting will be changed and
cycles in the following order. Upon changing the setting, the new setting is
briefly displayed as shown in “ ”. normal environment ( “norm E” ) → very
stable environment, ( “StAbL E” ) → unstable environment, ( “UnStb E” )
→ normal environment, ( “norm E” ). During use, the current environmental
setting of Pouring mode can be confirmed by observing the position of the
◀ symbol in the display.
(Display)
◀
When “StAbL E” is selected
◀
When “norm E” is selected
◀
When “UnStb E” is selected