10
0
15.3.3 Format
Set the balance output data format.
The standard format for the Shimadzu electronic balance: Select menu item 77.
The old output format for the Shimadzu electronic balance: Select menu item 78.
The old output format is employed in the following models.
EB-500, 5000, 280, 2800, AEL-1600, EB-50K (except -15)
In this format, the number of the lowest place of menu item 70 is assigned to identify
the balance.
Compatible format for the PR and SR series of Mettler-Toledo electronic balances.
Available commands, functions and responses are limited: Select menu item 79.
Compatible format for the IS series of Sartorius electronic balances. Available commands,
functions and responses are limited: Select menu item 80.
15.3.4 Communication Speed
Select the communication speed (300, 600, 1200, 2400, 4800, 9600, 19200, or 38400 bps).
Number of “b-xxx” shows bps (bits/second). Baud rate and bps are the same value.
Select one of the menu items 81 to 88 .
15.3.5 Parity / Bit Length
Select the parity and bit length.
No parity, 8-bit length: Select menu item 89.
Odd number parity, 7-bit length: Select menu item 90.
Even number parity, 7-bit length: Select menu item 91.
15.3.6 Stop Bit
Select the number of stop bits.
Stop bit 1: Select menu item 92.
Stop bit 2: Select menu item 93.