12.2 Percentage (%) Conversion
1. Set the % unit with menu item 56 if it is not set up.
The % unit is set before shipment.
2. Press the [UNIT] key several times in the mass
display until the % unit is displayed.
Setting the 100% reference
1. Press the [O/T] key to tare the balance.
2. Load the reference sample that corresponds to
the 100% value. This value must be equivalent to
100 counts or more in the “g” unit.
3. When the stability mark illuminates, press the
[CAL] key.“SEt” is displayed briefly and the
reference sample weight is displayed as 100%.
The weights of subsequent samples are displayed
as a percentage of the reference sample weight.
The numbers of digits displayed in the % unit and rounding off of the minimum digit
vary depending on the mass value of the reference sample and the balance model.
It is not possible to obtain resolution greater than that in the “g” unit.