66
13.2 Piece Counting (PCS)
1
. Set up the PCS with menu item 57 if it is not set.
(The PCS unit is set before shipment.)
2. Press the [UNIT] key several times in the mass
display until the PCS is displayed.
3. Load the container and press the [O/T] key to
tare the balance.
4. Count exactly five pieces (or 10, 20, 50, 100, or
200 pieces) of sample to be measured and load
them on the pan.
(Example)
5. Press the [CAL] key.
6. Every time the [CAL] key is pressed, the display
sequentially changes a
“Ld 5pcs” ....“Ld 200pcs”,
“Ld 5pcs”... The default setting is “Ld 10pcs”.
[O/T] key when the display is equivalent
eces.
Example: If 50 pieces are loaded, press the [O/T]
key when “Ld 50pcs” is displayed. This
determines the unit weight or average
weight per piece.
As sample is added or removed, the
piece count (number of pieces) is displayed.
Repeat steps 3 through 6 above when the sample or manufacturing lot is changed.
In this example, when the [CAL] key is pressed in the next PCS menu, display
starts from “Ld 50pcs”.