EasyManuals Logo

CAS CUX Series User Manual

CAS CUX Series
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
6
4
13.1.1 Checkweighing (Comparator) Display Type 1
This is the best mode to determine pass or failure judgment based on the sample weight.
Displays in use
1. Select menu map item 15 .
2. Set the upper threshold value, which corresponds
to the upper triangle mark, with menu item 16 .
3. Set the lower threshold value, which corresponds
to the lower triangle mark, with menu item 17 .
The decimal point is invisible in each value setting display. Setting the value cannot be
made in gram unit. Determine the number based on the minimum display of the
balance. For example, if the balance’s minimum display is 0.01g, 150g (150.00g in this
balance) must be set by inputting15000” in the display, not “150” or “150.00”.
Determination as follows:
Upper threshold < Sample weight HI
Lower thresholdSample weight Upper threshold GO
Sample weight < Lower threshold LO
13.1.2 Checkweighing (Comparator) Display Type 2
Use this mode for classification based on the sample weight. The display looks like a bar
graph, but also includes a checkweighing function.
Displays in use
1. Select menu map item 18 .
2. Set the upper threshold value, which corresponds
to the upper triangle mark, with menu item 19 .
3. Set the lower threshold value, which corresponds
to the lower triangle mark, with menu item 20 .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CAS CUX Series and is the answer not in the manual?

CAS CUX Series Specifications

General IconGeneral
BrandCAS
ModelCUX Series
CategoryScales
LanguageEnglish

Related product manuals