EasyManuals Logo

CAS CUX Series User Manual

CAS CUX Series
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
6
7
13.3 Auto Print
Auto Print function allows output of the data automatically without pressing the [PRINT]
key for each sample. The “Auto-Print symbol is illuminated when the Auto Print function
is activated.
Six types of Auto Print are selectable.
Print on loading: Select menu item 42 .
Load the sample when the value displayed is within the Zero Range. When stability is
detected and the positive displayed value is more than 5 times the Zero Range, data is
automatically output.
The next data output is not performed unless the display has returned to a value within
the Zero Range by unloading the sample or pressing the [O/T] key.
Print on loading and unloading: Select menu item 43 .
Load or unload the sample when the displayed value is within the Zero Range.
When stability is detected and the displayed positive or negative value is more than 5
times the Zero Range, data is automatically output.
The next data output is not performed unless the display has returned to a value within
the
Zero Range by unloading the sample or pressing the [O/T] key.
Print on loading and on zero: Select menu item 44 .
Load the sample when the value displayed is within the Zero Range. When stability is
detected and the positive displayed value is more than 5 times the Zero Range, data is
automatically output.
Unload the sample or press the [O/T] key. When the displayed value is within the Zero
Range and stability is detected, data is output again.
Print on loading, unloading, and on zero: Select menu item 45 .
Load the sample when the value displayed is within the Zero Range. When stability is
detected and the displayed positive or negative value is more than 5 times the Zero
Range, data is automatically output.
Unload the sample or press the [O/T] key. When the displayed value is within the Zero
Range and stability is detected, data is output again.
Print continuously: Select menu item 46 .
By pressing the [PRINT] key while the Auto-Print symbol and Stand-by symbol are lit,
the Stand-by symbol goes out, the Communication symbol lights and the displayed
data is continuously output.
Continuous output stops temporarily when the [PRINT] key is pressed.