12. Units
12.1 Unit Display Set-up
The UW/UX series balance can display weighed results in various weighing units.
1. It is possible to display units other than “g”.
Press the [UNIT] key in mass display to
sequentially change the selected units.
(Example)
When menu item 58 (carat)
has been selected.
2. Before weighing, set the display units to be used.
Gram, %, and PCS (piece counting) are set up
before shipment. Set up the desired units by
selecting menu from 54 to 62 .
For details of PCS and specific gravity
measurement, refer to 13.3 and 14.1 to 14.2.
For the unit names, refer to the menu map.
x In the unit setting menu, the stability symbol is illuminated to indicate the currently
set units.
x Set or release the unit by pressing the [O/T] key when the unit is displayed.
Gram cannot be released.
User unit (menu item 62 ): A numeric value (multiplier) is multiplied by the
gram (g) weight of the unit.
Error Display
Display Overload: This display appears if the mass
display exceeds 7 digits due to the choice of unit.