6
8
During continuous output, the Communication symbol may appear to remain lit.
If the transfer speed of the data output is slow, the display may become unstable.
Increase the transfer speed as much as possible and set the handshake off
(menu item 73 ).
Print on “GO” judge of checkweighing: Select menu item 47 .
When the checkweighing function (See 13.1) is used and stability is detected in the
“GO” range, the data is output once.
The next data output is executed after the displayed value is within the Zero Range.
x See 13.5 for details of Zero Range.
x See 11.3 for details of stability detection.
x Pretaring Value (13.7) function cannot be used with Auto Print functions.
x Auto Print mode setting can be cleared without menu operation, by pressing the
[POWER] key for about three seconds.
x Not more than one of these seven functions can be set at a time. Setting the second
one with the menu will automatically clear the previously set function;
Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval Timer(14.4), Add-on
mode(14.5), Animal Weighing mode(14.6), Formulation mode(14.7).
x For automatic outputs at preset intervals, see “14.4 Interval Timer”.
x When minimum display has been switched to 10d (See 9.2), the optimized Zero
Range for 1d display may cause failure in the operation (e.g. Data is not automatically
output as sample is loaded). Select optimal Zero Range value again when minimum
display has been changed (See 13.5 for details).