86
x Set the Stability and Response to Pouring mode (menu item 23 , See 11.2)
if faster response is required.
x When the [POWER] key is pressed in the Formulation standby state, the power
standby state is initiated.
x Formulation Mode setting can be cleared without menu operation, by pressing the
[POWER] key for about three seconds.
x See 11.3 for details of stability detection.
x See 13.5 for details of Zero Range.
x Cannot be used together with Pretaring (13.7).
x Weighing unit (12.1) and minimum display (9.2) cannot be changed when
Formulation Mode is on.
x When Formulation mode is ON, fully-automatic span calibration by PSC (10.3.2) or
Clock-CAL (10.3.3) is not performed. Weight symbol keeps blinking when span
calibration is necessary. Span calibration with the built-in calibration weight or external
weights can be performed between sets of measurements (Formulation stand-by).
x Not more than one of these seven functions can be set at a time. Setting the second
one with the menu will automatically clear the previously set function;
Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval Timer(14.4), Add-on
mode(14.5), Animal Weighing mode(14.6), Formulation mode(14.7).