9
8
Restricted Items in the Multi-Connection Mode
x Multi-Connection mode is not designed for each balance to independently send the data.
This mode is for sampling the data by control of multiple balances with one PC.
This is not the function to support multiple balances for sending the data separately.
When multiple number of data and commands exist on the system at the same time, the
balance may not operate normally. For example, multiple data may exist simultaneously
on the system when using multiple balances in the Application Measurement mode such
as Auto Print or Continuous Output, or pressing the [PRINT] key. Normal operation cannot
be guaranteed.
x Peripheral instruments connected to the DATA I/O or IFB-102A connector, such as an
EP- 50 Electronic Printer, cannot be used.
x The communication formats can only be used with “Format EB type”. Delimiters cannot
be used with menu items 97 or 98 (setting for WindowsDirect).
x It will take an average 0.05×N seconds for the data from the personal computer to reach
balance unit N, even at setting of 38400 bps. Also, it will take almost the same time for the
data from the balance in the Nth position to reach the personal computer.
Cancellation of Multi-Connection mode
Multi-Connection mode cannot be canceled unless the power for all balances is turned OFF.