Operation
Copyright © 2015 Coda Octopus Products Ltd
67
F180R MOTION Sensor User and Reference Guide
Figure 51: Calibration Status
When the F180R System has reached a calibrated state, a dialog will be displayed reporting
that the calibration has completed. You should then save the calibration values to the F180R
System. By doing this you will be updating the latest calibration values stored within the
system and in turn these will become the default settings used by the F180R System when it is
initially switched on or reset. The dialog also gives you the options of saving the calibration to
a text file on the local drive or dismiss and close the dialog without saving the calibration to
either unit or disk.
Figure 52: Save Calibration
You will also be able to save the calibrated settings to the F180R System from the Live tab ->
Calibration - Save.
After the settings have been stored, you can power–off and power–on the F180R System as
necessary without affecting the calibration—the calibration settings are saved to non–volatile
memory.
It is recommended that you also save the settings locally to your PC.