F180R MOTION Sensor User and Reference Guide
NOTE: The Remote Lever Arm data is not used during calibration and can be entered
later if required.
If the settings are correct then you should upload them and restart the F180R System from the
MOTION Settings Wizard.
4.3.2 Calibration Procedure
Once the configuration has been saved to the F180R System and the F180R System restarted,
the system will commence the calibration routine automatically.
NOTE: The system should only be calibrated when the vessel is performing a sequence
of dynamic manoeuvres—circles, figures–of–eight, sudden stops, sudden starts, etc.
When you are performing circular manoeuvres it is helpful to occasionally change the
direction of rotation. We recommend to perform figures of eight for best possible
calibration.
Typically, the F180R System calibration on a small boat will take from 30 to 60 minutes of
active manoeuvring. If your calibration is taking an unduly long time, or if you are
experiencing other difficulties in completing a calibration, consult the Appendix for
troubleshooting tips.
NOTE: DO NOT carry out a calibration when the vessel is stationary (e.g. tied up
alongside); in this condition, the motion sensor responses are dominated by noise,
which can lead to erroneous calibration data being obtained. If you suspect that
calibration data is the result of calibration in a non–dynamic environment, carry out a
new calibration.
NOTE: Depending on the vessel size it is not always possible to perform dynamic
manoeuvres. In this situation, the minimum requirement is that the vessel is
underway and thus subject to a degree of dynamic motion.
The current calibration status is displayed in the System Status area of the main application
and further details can be viewed on the Live tab -> Calibration -Status when an Ethernet
connection is active.
MOTION Control reports when the system calibration is complete in the System Status area in
the Calibration Status dialog. This will report Calibrated when an acceptable, within
specification calibration has been completed.