2-20
1010PVNFM-3A
Section 2
The default pipe configuration (Fully Developed) wll cause the flowmeter to use the conven-
tional Reynolds Compensation Table when compensating for liquid flow profile behavior.
Application Data Pipe Configuration Fully Developed*
(Change to Upstream 1 Elbow
Piping) Dbl Elbow +
Dbl Elbow -
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Anomaly Diams xxxx (numeric entry)
ð
ô
ð
Pipe Configuration Menu Structure
* Default
ô
ô
ð
Controlotron Dual Path [1] Path 1
Disignate Pipe Shape Near Transducers
Liquid Class Water 20C/68F
UniMass Table Create/Edit Table
Temperature Range -40F to 250F
Pipe Configuration Fully Developed
Anomaly Diams 1 Elbow
Dbl Elbow +
Dbl Elbow -
Valve
Expander
Reducer
Norm Entry
Header Inlet
Intrusions
Application Data
Use this menu cell to enter the
Use this menu cell to select
the pipe configuration that
most accurately represents
the upstream pipe condition.
PIPE CONFIGURATION OPTION LIST DEFINITIONS
Fully Developed Fully Developed flow, as would be expected for very long straight pipe
runs or installation downstream of a flow condition.
1 Elbow Single 90 degree Elbow upstream of transducer installation.
Dbl Elbow + Double out-of-plane Elbows upstream of transducer installation.
Dble Elbow - Double in-plane Elbows upstream of transducer installation.
Valve To Be Determined.
Expander Pipe expansion upstream of transducer installation.
Reducer Pipe reduction upstream of transducer installation.
Norm Entry To Be Determined.
Header Inlet Header or pipe manifold upstream of transducer installation.
Intrusions To Be Determined.