EasyManua.ls Logo

Controlotron 1010PVN - Kc Calibration; Kc Factor; Selecting the Kc Factor

Controlotron 1010PVN
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-50
1010PVNFM-3A
Section 2
Kc =
x 100
Actual Rate
Indicated Rate
-1
Kc Calibration
For most applications, the measured flow range produces a linear meter response. Therefore,
the Kc (slope correction) calibration is the preferred method since it only requires a single
correction factor for all the flow rates encountered.
Data Span/Set/Cal Calibrate Flowrate Intrinsic
Kc x.xxx
Calib. Table 1 Index Variable Flow
Vs
Liquident
Viscosity (cS)
Valc
Vaer
Temp 1
Reynolds #
Pressure
Aux
Calib. Table 1 New Point Cal. Index Value x.xxx
PosFlow Corr x.xxx
NegFlow Corr x.xxx
Accept/Clear Accept Pt.
Clear Pt.
Table Active 1 No
Clear Table 1 No
Yes
Data Span/Set/Cal Menu (Calibration Section)
ô
ð ð
ð
ð
ðð
ð
ô
ô
ô
ô
ô
ô
ð
ô
ô
ô
NOTE: Changing the calibration can cause profound changes in a flowmeter’s operat-
ing characteristics. Use only the most respected flow standard to obtain a cor-
rection factor. The percentage you enter must provide an accurate and consis-
tent shift across the entire flow range anticipated for the application.
Kc Factor
To obtain the Kc factor, compare flow total data taken simultaneously from the 1010 and a
reference meter whose accuracy meets the required standard. Allow both meters to accumu-
late flow total data long enough to average out any differences due to flow fluctuation between
the two meter locations. Compare outputs of the two totalizers to determine percentage in-
crease (+) or decrease (-) that is necessary to produce the best average correlation between
System 1010 and the reference standard.
Selecting the Kc Factor
When the [Kc] menu cell is selected, the meter imposes this % slope adjustment of its rate
output. Output data is zeroed and corrected for the Reynolds number (flow profile compen-
sated), however, in addition, a percent change in the rate output is imposed based on the data
entered in this cell. The number entered by the user is evaluated into a slope correction factor
by dividing it by 100 and algebraically adding it to 1. The resulting factor is used as a multi-
plier on the rate register of the instrument. Thus an entry of -3% will multiply the rate regis-
ter by 0.97, for example.
To calculate Kc:

Table of Contents

Related product manuals