SCU-1601 Control Unit for Turbomolecular Pump
*1 The driver output stops and the rotor continues rotating by inertia.
It may take
several hours to stop. To stop the STP pump quickly, close the vacuum valve at the outlet port
flange and introduce gas from the purge port into the STP pump.
Table
43 - Error
messages after self test (continued)
Overhaul is needed.
Contact Service office.
1. Contact Service office.
2. Contact Service office.
Overhaul is needed.
Contact
Service office.
Set "Warning Damage Point"
to "DISABLE" to release the
error and continue operation
temporary.
Frequency of "Disturbance"
or
"Mains
Failure
" error.
1. STP pump failure.
2. Failure of the SCU
-1601.
Frequency of "Disturbance"
or
"Mains
Failure
" error.
Failure of record in
memory
.
Failure of serial
communication
.
Abnormal board
connection
.
When the extent of the
damage on the touch
down bearing exceeds the
second damage limit.
Abnormal rotational speed
detection.
"Second Damage Limit"
occurs.