SCU-1601 Control Unit for Turbomolecular Pump
MT-96E-1B1-0
Page 149
8
MAINTENANCE
Referred
section
3.6.2
3.11.3
Table
44 - Error
messages after self test (continued)
Countermeasures
1. Connect the cable
correctly. If the cable
connector is not locked,
lock it securely.
2. Connect the cable
correctly. If the cable is not
locked, lock it securely.
3. Power ON the external
TMS unit and perform the
reset operation.
4. Contact Service office.
5. When the TMS function is
not in use, set the TMS
function to "DISABLE"
.
Probable
causes
1. Disconnection of the TMS
connection cable.
2. External TMS unit is not
connected
3. External TMS unit is not
powered ON
4. Failure of the TMS
connection cable.
5. Incorrect setting of the
TMS function.
Pump operation
Decelerate and stop.
Occurrence condition
TMS connection cable is
not connected when the
state "TMS Function" is
set to "ENABLE",
"External" or "Advanced"
.
LCD Messages
T. Cable
Disconnect
ed