STP-iS1607 Series Turbomolecular Pump
SERIAL COMMUNICATION PROTOCOL
Then the preceding SIM->PC character is "Ack", the instructed control command is executed and the
SIM returns the following response.
The PC transmits "Ack" or "Nak"; then transmits the next command if necessary.
Transmission frame when data is transmitted to two blocks (message is more than 256 characters
and less than 512 characters):
Designate the control command (the 1st block) on the PC
Next, the preceding SIM->PC character is "Ack", the PC continues instructing the control command
(the 2nd block).
Always assign less than 510 characters (n< 510) to the parameter so that the message is less than
512 characters.
Then the preceding SIM->PC character is "Ack", the instructed control command is executed and the
SIM returns the following response.
The PC transmits "Ack" or "Nak"; then transmits the next command if necessary.
5.3.4 Query command (in the RS232/RS485 single point connection)
A query command is used to read the pump operation state and setting values. The first character of
the query command in the RS232/RS485 single point connection is "?"(HEX code "3F"), and
succeeding characters are ASCII characters corresponding to the respective function code and
parameter.
CHR: Function code character, C
1
to C
n
: Parameter