TRANSIENT3000
104/162
3. Settings at the oscilloscope
Time base 10 to 50 ms,
Vertical deflection 5 V / division
On the CRO screen, the Burst must be visible. The single spike is not visible because the bandwidth is
insufficient.
7.4.2 ESD
1. Select 8 kV charging voltage and repetitions frequency 1Hz
2. Discharge to a ground plate. A spark of approximately 3 mm length must be visible.
7.4.3 SURGE
Verification as specified in the Basic Standard 1000-4-5.
Measurement of output voltage at no load
Measurement of short circuit current with short circuit output
Check that voltage and current waveforms are within the tolerances.
Calculate the source impedance from the peak voltage divided by the peak current.
1. Setting SURGE Test „Main Menu“
V = 1000 V; repetition 5s; coupling path L-N,
Attention!! The power cord must be removed from the inputs EUT Power 1 and 2 of the rear side.
2. Measuring points:
SURGE U-CRO for the voltage measurement at no load
SURGE I-CRO for current measurement at short circuit (make a short circuit on the front panel of the
TRA3000 F-S-D-V-C using a banana plug type cable 1000 between L-N)
3. Setting measuring equipment
Time base 5 µs,
Vertical deflection 0.5 V / division
Definition of the wave-forms and their tolerances, see Chapter 14.1
7.4.4 Interruption
Verification as specified in the Basic Standard 61000-4-11.
Trigger the measuring equipment via the external trigger input. Different trigger level, see Chapter 1.2.7
7.4.5 Variation
1. Setting TRA3000 F-S-D-V-C Setup Var 2s1s2s
2. Measuring point:
BNC output EUT Power U.
3. Setting measuring equipment
Time base 10 to 50 ms,
Vertical deflection 2 V / division
The voltage variation can be measured with the oscilloscope.