TRANSIENT3000
108/162
8.1.2 Error caused by running problem „Generator stopped“
EUT FAILED: Vpk: xxxxV > xxxxV
The selected voltage limits have been exceeded during SURGE
testing.
-Check limits
-EUT is defective.
EUT FAILED: Vpk: xxxkV <xxxkV
During SURGE test, the voltage has fallen below the selected
voltage limits:
-Check limits
-EUT is defective.
EUT FAILED: Ipk: xxxkA > xxxkA
The selected current limits have been exceeded during SURGE test.
-Check limits
-EUT is defective.
EUT FAILED: Ipk: xxxkA <xxxkA
During SURGE test, the current has fallen below the selected limits:
-Check limits
-EUT is defective.
EUT FAILED: External event
The input EUT failed has been activated ( grounded).
-Check EUT failed
-EUT is defective
EUT FAILED: by operator The operator has pressed FAIL on the front panel. .
Overcurrent: I-power : xxxA (>xxA) FOR generators with built in CDNs. The continuous current of the
EUT limit has been exceeded (AC)
Manual Trigger Timeout (>100sec) During SURGE and with manual trigger, the high voltage will be
switched off after 100 seconds, if no pulses have been released..
HV discharge timeout (>10s) FOR generators with EXT-TRA3000 E (ESD) the high voltage will be
switched off after 10 seconds, if no pulses have been released in air
discharge mode.