2 Safety
The TRA3000 F-S-D-V-C belongs to Safety class 1
2.1 Safety standard
The TRA3000 F-S-D-V-C fulfils the requirements of the safety standards IEC 61010 for laboratory
measurements equipment „Safety requirements for electrical measuring, control and laboratory equipment“.
Based on EN 61010 the declaration of conformity to low voltage directive (LVD 73/23/EEC O.J.N° L77,
1973-03-26) is given.
This manual is an integral part of the TRA3000 F-S-D-V-C tester. The instructions contained in the
manual regarding operation and the test set up are to be strictly observed.
2.2 Climatic Conditions
The TRA3000 F-S-D-V-C contains high voltage circuits in integrated form. EMC PARTNER only
guarantees a correct functioning of the tester TRA3000 F-S-D-V-C and the associated accessories, if the
TRA3000 F-S-D-V-C is operated in the climatic conditions specified.
Temperature 15 °C to 35 °C
Relative humidity 45 % to 75 %
Atmospheric pressure 86 kPa to 106 kPa (860 to 1060 mbar)
Not influenced by: direct solar radiation, rain or condensation water, dust or larger
electro magnetic fields as specified in the EMC compatibility chapter.
The TRA3000 F-S-D-V-C should be operated in a dry, clean room. If for any reason condensation water is
present in the TRA3000 F-S-D-V-C, then no TRA3000 operation should be started before the tester is
thoroughly dry.
It is strictly forbidden to operate the TRA3000 F-S-D-V-C in rooms with a gas explosion risk. The
high voltage of the TRA3000 F-S-D-V-C can generate sparks, which could ignite the gas.
People with heart pacemakers should not be in the vicinity of the test set up during operation.