EMC-PARTNER TRA3000 S
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TRANSIENT3000
24/162
With internal Variac:
Types of
loads:
switching
from to
Variable power
consumption
maximum
2.6 kW at UT 230 V.
With reduction of the voltage
the current is also reduced.
Examples: Ohmic -, inductive -,
capacitive -, mixed loads
Constant power consumption
maximum
1,2 kW at UT = 220V.
With reduction of the voltage the
current is increased.
Example: switched power supply
voltage change in
% of UT
at current
change 0 to 100 %
UT= voltage at EUT
Power 1
UT
100 %
100%
100%
100%
% UT
0 %
80%
70%
40%
current range r.m.s
0 to 16A
0 to 10 A
0 to 9 A
0 to 5 A
current range r.m.s
0 to 16A
0 to 5A
0 to 6 A
0 to 10 A
% of UT
0.7 %
4%
4%
5%
Note: all values apply for switching time at %UT< 5 s
1.3.7 Interruption and Voltage Variation IEC 61000-4-11 Ed.2 with external Variac
Types of
loads:
switching
from to
Variable power
consumption
maximum
3.7 kW at UT 230 V.
With reduction of the voltage
the current is also reduced.
Examples: Ohmic -, inductive -,
capacitive -, mixed loads
Constant power consumption
maximum
3,7 kW at UT = 220V.
With reduction of the voltage the
current is increased.
Example: switched power supply
voltage change in
% of UT
at current
change 0 to 100 %
UT= voltage at EUT
Power 1
UT
100 %
100%
100%
100%
% UT
0 %
80%
70%
40%
current range r.m.s
0 to 16A
0 to 12.8 A
0 to 11.2 A
0 to 6.5 A
current range r.m.s
0 to 16A
0 to 20A
0 to 23 A
0 to 40 A
% of UT
0.7 %
4%
4%
5%
Note: all values apply for switching time at %UT< 5 s
1.3.8 DIPS circuit in accordance with IEC 61000-4-29 for d.c. power ports.
Voltage range d.c. 20 to max. 300 V EUT Power
Current range 0 up to 16A 10A at 300V See derating curve of PS3
Inrush current capability at 110 V 220A Peak - 0%, +30% See 6.1.1
Interruption time 1ms up to 29999 ms
Rise and fall time at 100 Ohm load between 1 µs and 50
µs
See 6.1
IEC 61000-4-29 page 19:
The use of a generator with higher or lower voltage/current capability is allowed provided that the other
specifications are preserved. The test generator steady state power/current capability shall be at least 20%
greater than the EUT power/current ratings.