11BRemote Ports
130/162
Command TRIG
(TRIGger)
Explanation:
Set or query
Trigger Mode.
Argument: Characters AUTO, MAN
Example: TRIG MAN
TRIG?
Answer: MAN
Command SYF (SYncro Frequency)
Explanation: Set or query Syncro Frequency ( fundamental frequency).
Argument: Characters F1 correspond 16 Hz
F2 corresponds 40 Hz
F3 corresponds 50 Hz
F4 corresponds 60 Hz
F5 corresponds 400 Hz
Example: SYM ON
SYF F3
SYA 180
Command SYA (SYncro Angle)
Explanation: Set or query Syncro Angle [in degrees].
Argument:
Integer 0..360
Example: SYM
SYA 180
Command DEF (DEFaults)
Explanation: All parameter will be resetted to the default values.
Argument: no argument
Command PON (Power ON)
Explanation:
Turn on/off the
EUT power, or query the condition of the EUT power e. g. voltage value. These command
is only useful with automatic external CDN.
Argument: Integer 0,1,2
Example: (50Hz)
PON 1 (turn on the EUT power on PWR1)
PON?
Answer: 1
PON 0F (turn off the EUT power)