TRANSIENT3000
76/162
6.2.2.5 Synchronisation Power
Different possibilities exist to synchronize the
TRA-generator with a three phase coupling filter.
Example: TRA2000/TRA3000
The SURGE impulse must be synchronised with
the phase L1 to PE.
Set the black banana plugged cable for
Synchronisation and the yellow and red cable for
the surge pulse as shown in the pictures .
Cable for the synchronisation. Please use the 1
m black cable delivered with the CDN2000 to
synchronize the TRA3000 F-S-D-V-C with the
desired phase of the CDN2000.
In the pictures the synchronisation is made for
the Phase L1.
Additionally the neutral and the protective earth
must be connected on the rear side of between
the TRA3000 F-S-D-V-C and the CDN2000
First the synchronisation source must be
selected
When the power source is selected
Depending on the SURGE coupling (cable
yellow and red) the synchronisation angle must
be set as defined below:
Coupling: Phases to PE or N and synchronisation
(black cable on L1)
SURGE L1: 0° equal 0°
SURGE L2: 0° equal 120°
SURGE L3: 0° equal 240°
Coupling:
Phases to Phases and
synchronisation (black cable on L1)
SURGE L1 to L2: 0° equal 330°
SURGE L2 to L3: 0° equal 90°
SURGE L1 to L3: 0° equal 30°
Note: The 0° is defined as the angle at which the Surge is superimposed on the power supply depending
on the manually selected red yellow connections.