Emerson Process Management GmbH & Co. OHG6-80
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6.2.3 Setup Menu
6.2.3.6 Setup Alarms
Setup..
Alarms..
Alarms
Alarms Output Failsafe: Yes
Concentration..
Conc Average..
Temperature..
Pressure..
Flow..
Alarms 2
Calculator A..
Calculator B..
Calculator C..
Calculator D..
This menu and its submenus enable to con-
gure a couple of alarm conditions.
In case an alarm goes off, a status is set and
the related pictogram shows in the display.
Also digital outputs may be congured to be
activated (
6-68 ).
Congure concentration alarms ( 6-81 )
Congure alarms for concentration averages
(
6-83 )
Congure temperature alarms (
6-85 )
Congure pressure alarms (
6-85 )
Congure ow alarms (
6-85 )
X-STREAM XE analyzers can be upgraded
with optional software packages, to provide
four calculators (A ... D).
Their results can be monitored to activate
alarms (
6-86 )
.
If the system is setup accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Select if the alarms outputs are failsafe (Yes)
or not (No): Failsafe means, relay output coils
are powered during normal operation.