multi EA 5000 Troubleshooting
105
Error
code
Error message/cause Remedy
230 ABD: No connection
Communication error after ABD was detected dur-
ing initialization
Acknowledge message
Initialize
231 ABD: Runtime exceeded
Completion message for the ABD movement
takes too long
Check that the flame sensor is properly applied
and connected
Record log files
Inform customer service department
232 Flame sensor error
Flame sensor calibration failed Record log files
Inform customer service department
250 LPG: No connection
Communication interrupted after LPG detection
during initialization
Acknowledge message
Initialize
251 LPG: Runtime exceeded
Communication error
Completion message at dosing end not received
Acknowledge message
Check argon gas inlet pressure
Initialize
252 LPG: Argon missing during dosing
No argon applied to LPG module Check gas supply
Check gas inlet pressure
253 LPG: Incorrect sample volume
Dosed volume is not a whole multiple of the sam-
ple loop inserted
Adapt dosing volume to sample loop volume
260 Sample handling missing
No sampler module detected Connect at least one sampler module
Initialize
270 Auto-injector syringe: No connection
(applies to Autoinjector
, Autoinjector AI-EA)
No communication with auto-injector Acknowledge message
Initialize
271 Auto-injector syringe: Runtime exceeded
(applies to Autoinjector
, Autoinjector AI-EA)
Communication error
Completion message at dosing end not received
Acknowledge message
Check auto-injector
Initialize
272 Auto-injector syringe: Incorrect syringe size
Autoinjector: Dosing volume and syringe size dif-
ferent
Adapt the dosing volume and/or the syringe size
Autoinjector AI-EA cannot complete this order Initialize
273 Auto-injector syringe: Syringe drawn incorrectly
(Autoinjector
only)